Advanced Search
XIE Hong, QIU Xiang-jun, FU Lin, . Improvement of RT-HPLC on determination of lovastatin[J]. Chinese Journal of Public Health, 2003, 19(7): 814-815.
Citation: XIE Hong, QIU Xiang-jun, FU Lin, . Improvement of RT-HPLC on determination of lovastatin[J]. Chinese Journal of Public Health, 2003, 19(7): 814-815.

Improvement of RT-HPLC on determination of lovastatin

  •   Objective   To improve the detection concentration and time of plasma lovastatin by studying the RP-HPLC method.
      Methods   The condition of HPLC: The analytical column was packed with YWG-C18(10μm).The mobile phase was methanol-water(83:17, V/V).The temperature of column was 25℃.The compounds were quantified using a ultraviolet detector operated at 237 nm.The lovastatin in plasma sample extracted by cycllohexane-isopropyl alcohol.
      Results   The liner range of plasma lovastatin was 2.0-80 ng/ml.The regression recation was y=241.56+197.25x(r=0.999 1).The limit of detection was 2.0μg/L and the mean recovery was(96.01±7.17)%.The determination of a sample only needed 6 min.
      Conclusion   Controlling the pre2 treatment of plasma and the ratio of methanol to water was the way to enhance the detection limit and shorten detection time.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return